FDA Medical Device Safety
Infusion Pump Battery Correction: ICU Medical Updates Instructions for Use Regarding Batteries in Plum 360, A+ and A+3 Infusion System due to Diminished Battery Life that May Impact Infusion Delivery |
|
KCER Release Date: August 20, 2024 To: KCER Distribution list- including ESRD Network EDs and QIDs This recall involves updating instructions for using certain devices, and does not involve removing them from where they are used or sold. The FDA has identified this recall as the most serious type. This device may cause serious injury or death if you continue to use it without following the updated instructions. Affected Product
The FDA will continue to keep health care providers and the public informed if new or additional information becomes available. What to Do There is no need to return or discontinue use of Plum 360 or Plum A+ pumps. For clinical users:
On May 7, 2024, ICU Medical sent all affected customers an Updated Urgent Medical Device Correction letter recommending the following additional actions: Biomedical engineering:
Distributors:
Reason for Updates to Use Instructions ICU Medical is updating the use instructions for the Plum 360, Plum A+ and Plum A+3 Infusion Systems due to a manufacturing defect of the batteries, which can result in the life of the battery to substantially diminish. This updated notice expands affected products from a March 2023 notice and now includes all replacement batteries manufactured by CSB. The use of affected products may cause serious injury or death to patients due to interruption, under-infusion, or delays in the delivery of critical fluids, blood products, and medications. There has been one reported injury. There have been no reports of death. Device Use The Plum 360, Plum A+ and Plum A+3 Infusion Systems are large volume infusion pumps used to give fluids to patients in precisely controlled infusion amounts and rates. The system delivers blood or blood products, drugs, and other fluid mixtures through subcutaneous, intramuscular, intravenous, and intrathecal administration. The batteries and replacement batteries are used when the pump is not plugged into AC power, for example when a patient is being transported. Contact Information Customers in the U.S. should report adverse events or product complaints to Global Complaint Management: 1-844-654-7780 or ProductComplaintsPP@icumed.com. For additional information or other assistance, contact Technical Assistance, 1-800-241-4002, option 3. Additional FDA Resources
Unique Device Identifier (UDI) The unique device identifier (UDI) helps identify individual medical devices sold in the United States from manufacturing through distribution to patient use. The UDI allows for more accurate reporting, reviewing, and analyzing of adverse event reports so that devices can be identified, and problems potentially corrected more quickly.
How do I report a problem? Health care professionals and consumers may report adverse reactions or quality problems they experienced using these devices to MedWatch: The FDA Safety Information and Adverse Event Reporting Program.
|